KS4.3 T
0.1 - 4 mmol/l KS4.3
Acid / Indicator
20
S:4.3
Instrument specific information
The test can be performed on the following devices. In addition, the required cuvette
and the absorption range of the photometer are indicated.
Instrument Type Cuvette λMeasuring Range
MD 200, MD 600, MD 610,
MD 640, MultiDirect, PM 620,
PM 630
ø 24 mm 610 nm 0.1 - 4 mmol/l KS4.3
SpectroDirect, XD 7000,
XD 7500
ø 24 mm 615 nm 0.1 - 4 mmol/l KS4.3
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Alka-M-Photometer Tablet / 100 513210BT
Alka-M-Photometer Tablet / 250 513211BT
Application List
Waste Water Treatment
Drinking Water Treatment
Raw Water Treatment
Notes
1. The terms Alkalinity-m, m-Value, total alkalinity and Acid demand to KS4.3 are iden-
tical.
2. For accurate results, exactly 10 ml of water sample must be used for the test.
Display in the MD
100 / MD 110 /
MD 200
Bar code for the
detection of the
methods
Method name
Measuring range
Chemical Method
Method number
Language codes
ISO 639-1 Revision status
KS4.3 T / 20
How to use
ENMethod Reference Book4
EN
Implementation of the provision Acid capacity KS4.3 with Tablet
Select the method on the device
For this method, no ZERO measurements are to be carried out with the following de-
vices: XD 7000, XD 7500
10 ml
Fill 24 mm vial with 10 ml
sample.
Close vial(s).
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Dissolve tablet(s) by inver-
ting.
Sample
Place sample vial in the
sample chamber. • Pay
attention to the positioning.
Test
Press the TEST (XD:
START) button.
The result in Acid Capacity KS4.3 appears on the display.
Performing test
procedure
KS4.3 T / 20
NL Handboek van Methoden 01/20EN Handbook of Methods 01/20
How to use
ENMethod Reference Book 5
EN
Aluminium PP / M50
ENMethod Reference Book1.06
EN
Aluminium PP M50
0.01 - 0.25 mg/L Al AL
Eriochrom Cyanine R
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
VARIO Aluminium Reagent, Set F20 1 Pieces 535000
Preparation
1. To get accurate results the sample temperature must be between 20 °C and 25 °C.
2. To avoid errors caused by contamination, rinse the vial and the accessories
with Hydrochloric acid (approx. 20%) before the analysis. Then rinse them with
deionised water.
Aluminium PP / M50
ENMethod Reference Book1.0 7
EN
Determination of Aluminium with Vario Powder Pack
Select the method on the device.
Blank
20 mL
Prepare two clean 24 mm
vials. Mark one as a blank.
Put 20 mL sample in
100 mL measuring beaker
Add Vario ALUMINIUM ECR
F20 powder pack.
9
4
5
6
7
8
10
11 12
1
2
3
30 sec
Dissolve the powder by
mixing.
Press the ENTER button. Wait for 30 second(s)
reaction time.
Blank
1
Add Vario HEXAMINE
F20 powder pack.
Dissolve the powder by
mixing.
Place 1 drops Vario
ALUMINIUM ECR Masking
Reagent in the blank.
Aluminium PP / M50
ENMethod Reference Book1.08
EN
24 mm
10 mL
Place 10 mL pre-treated
sample in each vial.
Close vial(s). Press the ENTER button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Blank
Zero
Wait for 5 minute(s)
reaction time.
Place blank in the sample
chamber. Pay attention to
the positioning.
Press the ZERO button.
Sample
Test
Remove the vial from the
sample chamber.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
The result in mg/L Aluminium appears on the display.
Aluminium PP / M50
ENMethod Reference Book1.0 9
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Al 1
mg/l Al2O31.8894
Chemical Method
Eriochrom Cyanine R
Appendix
Interferences
Removeable Interferences
A low test result may be given in the presence of Fluorides and Polyphosphates. The
effect of this is generally insignificant unless the water has fluoride added artificially.
In this case, the following table should be used to determine the actual concentration
of aluminium.
Fluoride Displayed value: Aluminium [mg/L]
[mg/L
F]
0.05 0.10 0.15 0.20 0.25 0.30
0.2 0.05 0.11 0.16 0.21 0.27 0.32
0.4 0.06 0.11 0.17 0.23 0.28 0.34
0.6 0.06 0.12 0.18 0.24 0.30 0.37
0.8 0.06 0.13 0.20 0.26 0.32 0.40
1.0 0.07 0.13 0.21 0.28 0.36 0.45
1.5 0.09 0.20 0.29 0.37 0.48 ---
Bibliography
Richter, F. Fresenius, Zeitschrift f. anal. Chemie (1943) 126: 426
According to
APHA Method 3500-Al B
Chloride L (B) / M92
ENMethod Reference Book1.010
EN
Chloride L (B) M92
0.5 - 20 mg/L Cl-CL-
Mercury Thiocyanate / Iron Nitrate
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Chloride Reagent Set 1 Pieces 56R018490
Chloride L (B) / M92
ENMethod Reference Book1.0 11
EN
Determination of Chloride with liquid reagent
Select the method on the device.
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
20
Hold cuvettes vertically
and add equal drops by
pressing slowly.
Add 20 drops
KS251 (Chloride Reagenz
A).
Close vial(s).
Chloride L (B) / M92
ENMethod Reference Book1.012
EN
20
Invert several times to mix
the contents.
Add 20 drops
KS253 (Chloride Reagenz
B).
Close vial(s).
Sample
Test
Invert several times to mix
the contents.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
5 min
Wait for 5 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L Chloride appears on the display.
Chloride L (B) / M92
ENMethod Reference Book1.0 13
EN
Analyses
The following table identifies the output values can be converted into other citation
forms.
Unit Cite form Scale Factor
mg/l Cl-1
mg/l NaCl 1.65
Chemical Method
Mercury Thiocyanate / Iron Nitrate
Appendix
Interferences
Persistant Interferences
1. Reducing substances such as sulfite and thiosulfate, that can reduce iron (III) to
iron (II) or mercury (II) to mercury (I) may interfere. Cyanide, Iodine and Bromide
give a positive intereference.
Derived from
DIN 15682-D31
DIN ISO 15923-1 D49
Copper T / M150
ENMethod Reference Book1.014
EN
Copper T M150
0.05 - 5 mg/L Cua) Cu
Biquinoline
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
Copper No. 1 Tablet / 100 513550BT
Copper No. 1 Tablet / 250 513551BT
Copper No. 2 Tablet / 100 513560BT
Copper No. 2 Tablet / 250 513561BT
Set Copper No. 1/No. 2 100 Pc.#100 each 517691BT
Set Copper No. 1/No. 2 250 Pc.#250 each 517692BT
Preparation
1. Strong alkaline or acidic water samples must be adjusted to pH 4 to 6 before
analysis.
Copper T / M150
ENMethod Reference Book1.0 15
EN
Determination of Copper, free with tablet
Select the method on the device.
In addition, choose the test: free
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
Add COPPER No. 1 tablet
.
Crush tablet(s) by rotating
slightly.
Close vial(s).
Copper T / M150
ENMethod Reference Book1.016
EN
Sample
Test
Dissolve tablet(s) by
inverting.
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L free Copper appears on the display.
Determination of Copper, total with tablet
Select the method on the device.
In addition, choose the test: total
10 mL
Sample
Fill 24 mm vial with 10 mL
sample.
Close vial(s). Place sample vial in the
sample chamber. Pay
attention to the positioning.
Copper T / M150
ENMethod Reference Book1.0 17
EN
Zero
Press the ZERO button. Remove the vial from the
sample chamber.
24 mm
Add COPPER No. 1 tablet
.
Crush tablet(s) by rotating
slightly and dissolve.
Add COPPER No. 2 tablet .
24 mm
Crush tablet(s) by rotating
slightly.
Close vial(s). Dissolve tablet(s) by
inverting.
Copper T / M150
ENMethod Reference Book1.018
EN
Sample
Test
9
4
5
6
7
8
10
11 12
1
2
3
2 min
Place sample vial in the
sample chamber. Pay
attention to the positioning.
Press the TEST (XD:
START)button.
Wait for 2 minute(s)
reaction time.
Once the reaction period is finished, the measurement takes place automatically.
The result in mg/L total Copper appears on the display.
Copper T / M150
ENMethod Reference Book1.0 19
EN
Chemical Method
Biquinoline
Appendix
Interferences
Persistant Interferences
1. Cyanide CN- and Silver Ag+ interfere with the test result.
Method Validation
Limit of Detection 0.05 mg/L
Limit of Quantification 0.15 mg/L
End of Measuring Range 5 mg/L
Sensitivity 3.8 mg/L / Abs
Confidence Intervall 0.026 mg/L
Standard Deviation 0.011 mg/L
Variation Coefficient 0.42 %
Bibliography
Photometrische Analyse, Lange/Vedjelek, Verlag Chemie 1980
a) determination of free, combined and total | # including stirring rod, 10 cm
DEHA PP / M167
ENMethod Reference Book1.020
EN
DEHA PP M167
0.02 - 0.5 mg/L DEHA DEHA
PPST
Material
Required material (partly optional):
Reagents Packaging Unit Part Number
VARIO DEHA Reagent Set 1 Pieces 536000
Preparation
1. To avoid errors caused by iron deposits, rinse the glassware with Hydrochloric acid
(approx. 20%) before the analysis and then rinse with deionised water.
Remarks
1. Because the reaction depends on temperature, the temperature must be maintained
at 20 °C ± 2 °C.
2. Keep the sample vial in the dark or in the sample chamber during colour
development time. If the Reagent solution is exposed to UV-light (sunlight) it causes
high measurement results.
DEHA PP / M167
ENMethod Reference Book1.0 21
EN
Determination of DEHA (N,N-Diethylhydroxylamine) with Vario
Powder Pack and Fluid Reagent
Select the method on the device.
Blank
Blank
10 mL
Sample
10 mL
Prepare two clean 24 mm
vials. Mark one as a blank.
Put 10 mL deionised
water in the blank.
Put 10 mL sample in the
sample vial.
SampleBlank
Add a Vario OXYSCAV
1 Rgt powder pack in
each vial.
Close vial(s). Invert several times to mix
the contents.
SampleBlank
0.2 mL
Add 0.2 mL Vario DEHA
2 Rgt solution to each
vial.
Close vial(s). Invert several times to mix
the contents.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228
  • Page 229 229
  • Page 230 230
  • Page 231 231
  • Page 232 232
  • Page 233 233
  • Page 234 234
  • Page 235 235
  • Page 236 236
  • Page 237 237
  • Page 238 238
  • Page 239 239
  • Page 240 240
  • Page 241 241
  • Page 242 242
  • Page 243 243
  • Page 244 244
  • Page 245 245
  • Page 246 246
  • Page 247 247
  • Page 248 248
  • Page 249 249
  • Page 250 250
  • Page 251 251
  • Page 252 252
  • Page 253 253
  • Page 254 254
  • Page 255 255
  • Page 256 256
  • Page 257 257
  • Page 258 258
  • Page 259 259
  • Page 260 260
  • Page 261 261
  • Page 262 262
  • Page 263 263
  • Page 264 264
  • Page 265 265
  • Page 266 266
  • Page 267 267
  • Page 268 268
  • Page 269 269
  • Page 270 270
  • Page 271 271
  • Page 272 272
  • Page 273 273
  • Page 274 274
  • Page 275 275
  • Page 276 276
  • Page 277 277
  • Page 278 278
  • Page 279 279
  • Page 280 280
  • Page 281 281
  • Page 282 282
  • Page 283 283
  • Page 284 284
  • Page 285 285
  • Page 286 286
  • Page 287 287
  • Page 288 288
  • Page 289 289
  • Page 290 290
  • Page 291 291
  • Page 292 292
  • Page 293 293
  • Page 294 294
  • Page 295 295
  • Page 296 296
  • Page 297 297
  • Page 298 298
  • Page 299 299
  • Page 300 300
  • Page 301 301
  • Page 302 302
  • Page 303 303
  • Page 304 304
  • Page 305 305
  • Page 306 306
  • Page 307 307
  • Page 308 308
  • Page 309 309
  • Page 310 310
  • Page 311 311
  • Page 312 312
  • Page 313 313
  • Page 314 314
  • Page 315 315
  • Page 316 316
  • Page 317 317
  • Page 318 318
  • Page 319 319
  • Page 320 320
  • Page 321 321
  • Page 322 322
  • Page 323 323
  • Page 324 324
  • Page 325 325
  • Page 326 326
  • Page 327 327
  • Page 328 328
  • Page 329 329
  • Page 330 330
  • Page 331 331
  • Page 332 332
  • Page 333 333
  • Page 334 334
  • Page 335 335
  • Page 336 336
  • Page 337 337
  • Page 338 338
  • Page 339 339
  • Page 340 340
  • Page 341 341
  • Page 342 342
  • Page 343 343
  • Page 344 344
  • Page 345 345
  • Page 346 346
  • Page 347 347
  • Page 348 348
  • Page 349 349
  • Page 350 350
  • Page 351 351
  • Page 352 352
  • Page 353 353
  • Page 354 354
  • Page 355 355
  • Page 356 356
  • Page 357 357
  • Page 358 358
  • Page 359 359
  • Page 360 360
  • Page 361 361
  • Page 362 362
  • Page 363 363
  • Page 364 364
  • Page 365 365
  • Page 366 366
  • Page 367 367
  • Page 368 368
  • Page 369 369
  • Page 370 370
  • Page 371 371
  • Page 372 372
  • Page 373 373
  • Page 374 374
  • Page 375 375
  • Page 376 376
  • Page 377 377
  • Page 378 378
  • Page 379 379
  • Page 380 380
  • Page 381 381
  • Page 382 382
  • Page 383 383
  • Page 384 384
  • Page 385 385
  • Page 386 386
  • Page 387 387
  • Page 388 388
  • Page 389 389
  • Page 390 390
  • Page 391 391
  • Page 392 392
  • Page 393 393
  • Page 394 394
  • Page 395 395
  • Page 396 396
  • Page 397 397
  • Page 398 398
  • Page 399 399
  • Page 400 400
  • Page 401 401
  • Page 402 402
  • Page 403 403
  • Page 404 404
  • Page 405 405
  • Page 406 406
  • Page 407 407
  • Page 408 408
  • Page 409 409
  • Page 410 410
  • Page 411 411
  • Page 412 412
  • Page 413 413
  • Page 414 414
  • Page 415 415
  • Page 416 416
  • Page 417 417
  • Page 418 418
  • Page 419 419
  • Page 420 420
  • Page 421 421
  • Page 422 422
  • Page 423 423
  • Page 424 424
  • Page 425 425
  • Page 426 426
  • Page 427 427
  • Page 428 428
  • Page 429 429
  • Page 430 430
  • Page 431 431
  • Page 432 432
  • Page 433 433
  • Page 434 434
  • Page 435 435
  • Page 436 436
  • Page 437 437
  • Page 438 438
  • Page 439 439
  • Page 440 440
  • Page 441 441
  • Page 442 442
  • Page 443 443
  • Page 444 444
  • Page 445 445
  • Page 446 446
  • Page 447 447
  • Page 448 448
  • Page 449 449
  • Page 450 450
  • Page 451 451
  • Page 452 452
  • Page 453 453

Lovibond 00386480 - MD1x0/200 - Chloride L (B), Aluminium PP, Copper T, DEHA PP, Hydrazine P, Iron LR L (A), Oxygen dissolved C, Phosphate HR L, Polyacrylate L, Silicate HR PP Manual do usuário

Tipo
Manual do usuário